| Model Number |
38265 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned product consisted of an imager diagnostic catheter in the un-opened sterile pouch.Tensile strength inspection of the returned product was performed and the device failed to meet specification for tip detachment.The device was functionally tested with a test guidewire.The imager tip was detached during advancement of the guidewire.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
An unused 5f pigtail imager ii angiographic catheter in it's unopened original packaging was returned to boston scientific from a customer associated with mdr id 2134265-2019-13161.There was no reported complaint against this returned device.The device was intact upon receipt.Boston scientific carried out a tensile strength test on the returned device resulting in a tip detachment concluding that the device failed to meet specification.This event is associated with previously reported field action 92484513-fa with product being removed from the field.
|
| |
|