Model Number 19191
Device Problem Insufficient Information (3190)
Patient Problem Hyperglycemia (1905)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
A tear was found in the unexposed portion of the soft cannula.This effectively diverted fluid flow the tip of the soft cannula, resulting in the device being unable to effectively deliver insulin.No other problems were identified with the device.
 
Event Description
It was reported that the patient's blood glucose (bg) values reached 430 mg/dl, while wearing the pod between 24 and 36 hours on the abdomen.For treatment, the patient gave correction boluses, replaced the pod and drank water.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
jordan biswurm
100 nagog park
acton, MA 01720
9786007000
MDR Report Key10474999
MDR Text Key205505636
Report Number3004464228-2020-13831
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385081120033
UDI-Public(01)20385081120033(11)200402(17)211002(10)L71022
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/02/2021
Device Model Number19191
Device Catalogue NumberZXP425
Device Lot NumberL71022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received08/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age11 YR
-
-