| Model Number |
7010 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 3.0mm x 12mm quantum maverick balloon catheter was advanced for post-dilation.However, during inflation, the balloon ruptured.The procedure was completed with a different device.No known patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 3.0mm x 12mm quantum maverick balloon catheter was advanced for post-dilation.However, during inflation, the balloon ruptured.The procedure was completed with a different device.No known patient complications nor injuries reported.It was further reported that the 80% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery.The balloon ruptured during third inflation at 12 atmospheres for 5 seconds.There were no patient complications nor injuries reported and the patient's status was fine.
|
| |
|