• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and mildly calcified forearm shunt blood vessel.A 4.0mmx40mmx80cm sterling balloon catheter was advanced for dilatation.However, during first inflation at 14 atmospheres, the balloon ruptured.The procedure was completed with a different device.There were no patient complications nor injuries reported and the patient status was good.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There were no issues detected.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and mildly calcified forearm shunt blood vessel.A 4.0mmx40mmx80cm sterling balloon catheter was advanced for dilatation.However, during first inflation at 14 atmospheres, the balloon ruptured.The procedure was completed with a different device.There were no patient complications nor injuries reported and the patient status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10475026
Report Number2134265-2020-12089
Device Sequence Number2809276
Product Code DQY
UDI-Device Identifier08714729123637
UDI-Public08714729123637
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2023
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0025422132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Initial Date Received by Manufacturer 08/25/2020
Supplement Date Received by Manufacturer10/06/2020
Initial Report FDA Received Date08/31/2020
Supplement Report FDA Received Date10/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-