SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568S |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Event Description
|
It was reported that the power of the light source was down.No case reported´therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of power failure could not be confirmed.Product passed all functional testing.All functions perform as expected and no loss of power occurred during testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|