| Model Number |
3L80-32 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier: (b)(6).All available patient information is included.Additional patient details are not available.An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely depressed architect co2 results for four patients.The customer alleges having to issue about 500 corrected co2 reports.However, no data is available for this allegation.The customer did provide information for four patients: pt#1 (b)(6): initial = 16 mmol/l, repeat = 22 mmol/l; pt# 2: (b)(6)initial = 20 mmol/l, repeat = 27 mmol/l; pt#3: (b)(6): initial = 14 mmol/l, repeat = 25 mmol/l; pt#4: (b)(6): initial = 16 mmol/l, repeat = 25 mmol/l (normal range: 20-29 mmol/l).There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the customers issue included a review of tickets for carbon dioxide (co2) ln 3l80 lot 56883uq02, and trending review.The ticket review determined that there is normal complaint activity for this lot.The trending review determined no trend for falsely depressed results for the product.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Return testing was not completed as returns were not available.Data was provided fregarding the patient sample where the sample was retested by thecustimer, and acceptable results were produced.Based on all available information, no product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|