| Model Number |
86-6026 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The patient was revised to address loosening of the tibial component at cement to implant interface.An unknown cement was used.At primary surgery, surgeon only put cement on the under surface of the baseplate and none on the stem/keel portion.Sigma femur was well fixed.Doi: unknown, dor: (b)(6) 2020, unknown affected side.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
|
| |
|
Search Alerts/Recalls
|