| Model Number |
9545C |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that one of his blood glucose readings was not stored in the memory of the contour meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour meter and contour test strips for evaluation, however, the returned test strips were expired, therefore, the in-house contour test strips were tested with the returned meter using a blood sample, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
|
| |
|