| Model Number |
7252 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that her blood glucose readings from (b)(6) 2020 and (b)(6) 2020 were not stored in the memory of the contour next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the contour next ez meter with serial # (b)(6) for evaluation.However, the returned meter was different from the one implicated to be involved in the event i.E.Contour next ez meter with serial # (b)(6).No test strips were returned.Therefore, the returned meter was tested with the in-house contour next test strips using a blood sample, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly stored in the meter's memory.Additionally, a device history record was reviewed for the suspected contour next ez meter (serial # (b)(6)) and no manufacturing anomalies were found.
|
| |
|