• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 38 MM DIAMETER; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 38 MM DIAMETER; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Damage to Ligament(s) (1952); Pain (1994); Scar Tissue (2060); Swelling (2091); Tissue Damage (2104); Joint Dislocation (2374); Patient Problem/Medical Problem (2688)
Date of Event 08/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Therapy date : item#:42522401013; articular surface fixed bearing posterior stabilized (ps) right 13 mm height lot#:63133891.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
Bilateral tka performed approximately one year ago.Patient experienced a fall and several weeks later developed right knee pain, swelling, and patella was found to be subluxing/dislocating.Patient underwent revision of right knee patella and poly liner.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The product was evaluated through manufacturing review and the reported event was confirmed through review of medical records.The device history records were reviewed and no discrepancies were identified.As it is unknown what caused the fall and patellar dislocation, a definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL POLY PATELLA CEMENTED 38 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key10478427
Report Number0002648920-2020-00393
Device Sequence Number2991417
Product Code MBH
UDI-Device Identifier00889024247659
UDI-Public(01)00889024247659
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42540000038
Device Lot Number64422815
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/17/2020
Supplement Date Received by Manufacturer10/16/2020
Initial Report FDA Received Date09/01/2020
Supplement Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight116
-
-