| Model Number |
KIT, TMX NFRS STARTERMG/DL |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Meter was returned for evaluation; no defect was detected test strips were not returned for evaluation most likely underlying root cause: mlc-58: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer in a follow-up call on 11-aug-2020 to ensure the replacement products resolved the initial concern - able to establish contact with customer and stated replacement product resolved initial concern.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results obtained of 177, 190 and 165 mg/dl.The customer¿s expected fasting blood glucose test result range is 70-130 mg/dl.The customer felt well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a blood test was performed by the customer non-fasting and produced test result of 197 mg/dl using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer¿s expiration date is 10/31/2021 and open vial date is (b)(6) 2020.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|