Model Number N/A |
Device Problems
No Display/Image (1183); Failure to Pump (1502); Pumping Stopped (1503)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.A getinge field service engineer (fse) has advised that the iabp is undergoing testing, but the repair has not yet started because the source of the problem has not yet been found.Also, the fse has reported that the customer may decide not to repair unit do to age of the pump.A supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) stopped pumping and the screen went black.There was no alarm.The iabp did restart but was replaced.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) has advised that the iabp is permanently out of service.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) stopped pumping and the screen went black.There was no alarm.The iabp did restart but was replaced.No patient harm, serious injury or adverse event was reported.
|
|