MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LS3112
Medical Device Problem Codes Failure to Fire (2610); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The incident device has been received and is under evaluation.When the device evaluation is complete a follow-up report will be submitted.
 
Event or Problem Description
According to the reporter, during a procedure, the device did not activate.A similar device was used to complete the case.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a jaw detachment failure.Functional testing found no visual defects except broken plastic snaps.The device was attached to the re-usable metal clamp and connected to the generator.It was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactory and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.It was reported that the device did not activate when keyed.The reported issue could not be confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key10478893
Report Number1717344-2020-00979
Device Sequence Number3120224
Product Code GEI
UDI-Device Identifier10884524002514
UDI-Public10884524002514
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K981916
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2023
Device Model NumberLS3112
Device Catalogue NumberLS3112
Device Lot Number82910177X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer09/24/2020
Initial Report FDA Received Date09/01/2020
Supplement Report FDA Received Date01/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-