Model Number 931-36-750
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to an infection.The previous surgery and the surgery detailed in this event occurred 12 days apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available.The revision surgery was completed as intended and without incident.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.Nonconforming material report history below.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Ncmr# 53651 item: 931-36-750 lot 010v1152 lot quantity-10, sample quantity-0, reject quantity-10, failure code-o13 - gas plasma sterilization failed, cause code-eq01 - machine, revision found-a, defect description- power failure during sterility run e191220o causing gas plasma sterilization to fail, specification requirements-.2000.2140 cycle must be completed and pass, disposition-rework, justification- see attached rework steps to re-sterilize parts.Parts will not be eligible for further re-sterilization after rework is completed.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was reported as an infection.There were no findings during this evaluation that indicate that the reported device was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP EXTENDED LINERS
Common Device Name
LINER/NON-HOODED-NEU, MP6, HXE-PLUS, 36MM
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd
austin, tx
MDR Report Key10480097
Report Number1644408-2020-00754
Device Sequence Number2795620
Product Code OQG
UDI-Device Identifier00190446177913
UDI-Public(01)00190446177913
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/19/2024
Device Model Number931-36-750
Device Catalogue Number931-36-750
Device Lot Number010V1152
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2020
Initial Report FDA Received Date09/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
-
-