• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RND ENDBTN POST SST W-SZ3/4 SUTURE LOOP; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RND ENDBTN POST SST W-SZ3/4 SUTURE LOOP; WASHER, BOLT NUT Back to Search Results
Model Number 71934993
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); Joint Dislocation (2374); Osteolysis (2377)
Date of Event 08/04/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a glenoid bone loss surgery, a revision had to be performed.The surgeon noted excessive resorption on the distal tibial allograft associated with the use of the round endobutton, as compared to the resorption when using screw fixation.The resorption was noted in the 2 year follow-up mri and confirmed during the revision surgery.The patient's outcome is unknown.
 
Additional Manufacturer Narrative
Report was inadvertently submitted under manufacturer number ¿(b)(4)¿ but the correct manufacturer number is ¿(b)(4).¿ d3 and g1 information was updated.
 
Additional Manufacturer Narrative
Lot number, h6: updated.
 
Additional Manufacturer Narrative
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states the provided report indicates resorption of the distal tibial allograft noted on an mri taken at two years post op.And a x-ray image dated the same month as the primary shows 4 buttons and some radiolucency around the inferior one of this view.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H6: health effect - clinical code and health effect - impact code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RND ENDBTN POST SST W-SZ3/4 SUTURE LOOP
Common Device Name
WASHER, BOLT NUT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10480267
Report Number1020279-2020-04360
Device Sequence Number14110454
Product Code HTN
UDI-Device Identifier00885556424544
UDI-Public00885556424544
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K183232
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71934993
Device Catalogue Number71934993
Device Lot Number19BM01785
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer09/08/2021
09/08/2021
02/10/2023
Initial Report FDA Received Date09/01/2020
Supplement Report FDA Received Date09/08/2021
12/07/2021
02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-