Model Number DAQ916
Medical Device Problem Codes Use of Device Problem (1670); Device Sensing Problem (2917)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found no anomalies with the unit.The unit has been received in good condition.The unit has been successfully tested according to service repair instructions.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported that the preamp was broken.There was no patient impact.
 
Event or Problem Description
Additional information received mentioned that there was no stimulation during the procedure.They did not hear/see anything that made them aware that there was something wrong with the device.They tried switching the device off and on.
 
Additional Manufacturer Narrative
B3: additional information suggest that event date:2020-08-10 no longer apply for this event.H6: additional information suggest that fdd:c63318 no longer apply to this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE PREAMPLIFIER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key10481240
Report Number1045254-2020-00388
Device Sequence Number19489889
Product Code GWF
UDI-Device Identifier00613994910950
UDI-Public00613994910950
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDAQ916
Device Catalogue NumberDAQ916
Device Lot Number206687213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer10/15/2020
Initial Report FDA Received Date09/02/2020
Supplement Report FDA Received Date11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-