MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1200 SYNCHRONY PIN
Device Problems Device Appears to Trigger Rejection (1524); Migration (4003)
Patient Problems Failure of Implant (1924); No Information (3190)
Event Date 08/18/2020
Event Type  Injury  
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
There was an extrusion of the electrode out of the cochlea.The user will be explanted on the (b)(6).
 
Manufacturer Narrative
Conclusion: device investigations did not reveal any device defect or problem which is expected to have been present whilst implanted.This finding was expected, because according to the patient report the active electrode migrated postoperatively completely outside of the cochlea.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
 
Event Description
Migration of the electrode out of the cochlea was observed.The concerned device was explantation on (b)(6) 2020.The recipient received a device from another manufacturer on the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key10481317
MDR Text Key205267979
Report Number9710014-2020-00487
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737094058
UDI-Public(01)09008737094058
Combination Product (y/n)N
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
-
-