• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US BIOINTRAFIX 6-7MMX30MMTPRSCR; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US BIOINTRAFIX 6-7MMX30MMTPRSCR; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 254622
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter phone number (b)(6).Udi: (b)(4).
 
Event or Problem Description
It was reported that during the surgery of anterior cruciate ligament reconstruction, the anchor was broken off, removed all the pieces.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during the surgery of anterior cruciate ligament reconstruction, the anchor was broken off, removed all the pieces.The complaint device was received and inspected.The screw is broken near its distal end confirming this complaint.Visual observation indicates the external geometry of the screw was not damaged as the threads did not strip during insertion.The possible root cause for issue experienced can be attributed to an excessive torque applied during the insertion, whether the insertion was off axis into the bone hole.However, it cannot be conclusively affirmed. a manufacturing record evaluation was performed for the finished device lot number: 3l18858, and no non-conformance were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOINTRAFIX 6-7MMX30MMTPRSCR
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10481657
Report Number1221934-2020-02348
Device Sequence Number11766346
Product Code HWC
UDI-Device Identifier10886705013197
UDI-Public10886705013197
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number254622
Device Catalogue Number254622
Device Lot Number3L18858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer09/02/2020
09/04/2020
Initial Report FDA Received Date09/02/2020
Supplement Report FDA Received Date09/03/2020
09/08/2020
Patient Sequence Number1
-
-