OLYMPUS MEDICAL SYSTEMS CORP. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-5
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However there was the possibility that this phenomenon was attributed to a failure of the ccd or the electrical circuit board inside the video connector, or a malfunction of the video system center.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus service operation repair center (sorc) was informed from the facility that in unspecified timing the communication error between the subject device and video system center occurred.Sorc checked the subject device and found that the reported issue could not duplicated.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10481856
MDR Text Key224635776
Report Number8010047-2020-06101
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080948
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/06/2020
Initial Date FDA Received09/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-