| Model Number |
802512 |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Patient Problem/Medical Problem (2688)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient stated the catheter alignment was not correct, it would not insert right and was difficult to insert.Also stated in the past, patient had catheters where coude tip was not perfectly a 45 degree angle.Stated this caused patient bleeding.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed since the product meets specifications stating that "accept any curve unless catheter is completely straight".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "instructions for use indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands.Caution: federal (u.S.A.) laws restricts this device to sale by or on the order of a physician.".
|
| |
|
Event or Problem Description
|
|
It was reported that the patient stated the catheter alignment was not correct, it would not insert right and was difficult to insert.Also stated in the past, patient had catheters where coude tip was not perfectly a 45 degree angle.Stated this caused patient bleeding.No medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|