MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.0/26; COVERED CORONARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.0/26; COVERED CORONARY STENT Back to Search Results
Model Number 434899
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Death (1802)
Date of Event 07/31/2020
Type of Reportable Event Death
Event or Problem Description
A pk papyrus covered stent system was selected for treatment of a perforation which occurred one day before.During procedure it seemed that the stent was displaced on the delivery system and therefore the device was removed from patients body.There was no attempt to deploy the device and it had no effect on the procedure.Another pk papyrus was inserted.Patient passed away 2 hours after intervention.It was stated by the hospital staff that the death was not device related.
 
Additional Manufacturer Narrative
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 3.0/26
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10482834
Report Number1028232-2020-03725
Device Sequence Number3028269
Product Code NIV
UDI-Device Identifier07640130447271
UDI-Public07640130447271
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model Number434899
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03193056
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer11/03/2020
Initial Report FDA Received Date09/02/2020
Supplement Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-