| Model Number |
86640 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was reported for the amp not turning on.Visual inspection of the device revealed that it was received in good condition with no visible damage.During functional testing it was confirmed that the amplifier will not turn on, the led power did not light and the fan did not turn on.It was also revealed that 2 fuses had continuities.
|
| |
|
Event or Problem Description
|
|
It was reported hat the procedure was canceled.It was reported that a clearsign amp, 80 was selected for a procedure.It was noted that the amp did not turn on and the issue was unable to be resolved.It was not reported if the patient was sedated prior to this issue but no patient complications were reported.
|
| |
|
Search Alerts/Recalls
|