| Model Number |
367874 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use tubes were discovered to be collapsed with 800 bd vacutainer® sodium heparinn (nh) 158 usp units blood collection tubes.The following information was provided by the initial reporter: collapsed tubes.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for collapsed tubes with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Note that there are potentially varying degrees of tube collapse, based on the temperature and duration of time at elevated temperature.Mildly collapsed tubes will retain vacuum and only show minimal deformation.Mildly collapsed tubes will draw appropriately, but may not be able to be processed on laboratory instruments (e.G.May be slightly bowed and no longer fit in a rack).As long as the vacuum is retained and the tube fills appropriately, the tube will function properly.Fully collapsed tubes appear flattened, twisted, and visibly deformed.Fully collapsed tubes retain little or no vacuum and therefore cannot be drawn to an appreciable volume.These tubes are easily identified before use by the healthcare worker, so there is no potential impact to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use tubes were discovered to be collapsed with 800 bd vacutainer® sodium heparinn (nh) 158 usp units blood collection tubes.The following information was provided by the initial reporter: collapsed tubes.
|
| |
|
Search Alerts/Recalls
|