• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. VERIFY ENHANCED; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. VERIFY ENHANCED; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 353101
Medical Device Problem Codes Energy Output Problem (1431); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Therapeutic Effects, Unexpected (2099)
Date of Event 08/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical devices: product id: 306006, lot#: 60226826, product type: screening.Other relevant device(s) are: product id: 306006, serial/lot #: (b)(4), ubd: 10-feb-2022, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The manufacture representative reported that during the basic evaluation the electrode didn't work.The sphincter didn't contract during measurement.After several trials we decided to implant a new basic evaluation lead.Afterwards the sphincter contracted as it should.It was noted that there was no improvement of symptoms, because of a lack of stimulation at sacral nerves.It was noted that a new basic evaluation lead was implanted on (b)(6) 2020.
 
Additional Manufacturer Narrative
Continuation of d11: product id: 306006, lot#: 60226826, product type: screening device; product id: neu_ stylet_acc, product type: accessory.H3: the returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified that the conductors were stretched in the body of the lead; however, electrical testing determined that continuity was complete and there were no electrical shorts between the circuits.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIFY ENHANCED
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10483268
Report Number3004593495-2020-01001
Device Sequence Number2591049
Product Code EZW
UDI-Device Identifier00643169968783
UDI-Public00643169968783
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number353101
Device Catalogue Number353101
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2020
Supplement Date Received by Manufacturer10/27/2020
Initial Report FDA Received Date09/02/2020
Supplement Report FDA Received Date10/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Outcome Attributed to Adverse Event Required Intervention;
-
-