|
This report is being supplemented to provide the legal manufacturer¿s investigation summary regarding this report.From the information provided during complaint initiation, the following was confirmed: it was confirmed that the brush tip was stuck in the balloon channel, and that there was no description of the type of brush used.A review of the product labeling provides the following information: in regard to the brush, confirming the contents of the ifu which was provided during delivery of the product, the following description was found.All items mentioned below are single-use items.Do not clean, disinfect, or sterilize the equipment prior to use and/or after use.Doing so may damage the equipment.Using damaged equipment may make it difficult to effectively reprocess the endoscope, and could cause endoscope and/or equipment damage.Mh-507,bw-411b, and maj-1339.This is an event that occurred during reprocessing, not during a procedure.Although the true cause has not been identified because this event was confirmed based on the image and information alone, the following is presumed: the image shows that the brush tip has been stuck in the balloon channel and is broken.In brushing the balloon groove during reprocessing, the brush tip may have been stuck in the balloon channel.When the brush tip was stuck in the balloon channel, the brushing force may have been applied, resulting in the breakage of the brush.The type of brush used this time is unknown.If this event occurred using a brush other than the recommended brush (mh-507,bw-411b,maj-1339) described in the instruction manual, the safety and cleaning efficacy of bf-uc180f cannot be assured.
|