Model Number 16-02-85 |
Device Problem
Insufficient Heating (1287)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t was not heating to the set temperature during setup.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could reproduce the reported issue.He replaced both heating elements in the patient tank and solved the issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Event Description
|
See initial report.
|
|