| Model Number |
01.26.58MB |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed by medacta regulatory affairs department on 13.08.2020: lot 160207: (b)(4) items manufactured and released on 2-may-2016.Expiration date: 2021-04-18.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported event.Other device involved in the event: cup: versafitcup cc trio 01.26.45.1158 acetabular shell cc trio no-hole ø 58 lot.2000144 (k122911).Batch review performed by medacta regulatory affairs department on 13.08.2020: lot 2000144 : (b)(4) items manufactured and released on 30-apr-2020.Expiration date: 2025-04-20.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported event.
|
| |
|
Event or Problem Description
|
|
The surgeon wanted to implant a versafitcup dm s58 and after he had a press-fit the impacting ring did not come off.By trying several times afterward also with a versafitcup cc trio he had to cement a new one- apricot.The bone structure was damaged.The surgeon cemented a new cup on the bone damage.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection: it is possible that a sizing occurred due to an incorrect usage: - wrong screwing of the shell versaficup #58 dm ref.01.26.10.0139 with the impacting ring #58 ref.01.26.10.0139.
|
| |
|
Search Alerts/Recalls
|