• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364505
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2020
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was chosen for treatment of a severely calcified lesion (99 percent stenosis degree in severely tortuous mid lad.) the lesion could not be crossed with the device.It was detected that the stent was deformed.
 
Additional Manufacturer Narrative
Combination product: yes the returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is deformed in its center.Few struts are bent outwards.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the bent struts were initially crimped on the balloon.Outside of the deformed zone the crimped diameter of the stent still complies with the specification.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10488797
Report Number1028232-2020-03763
Device Sequence Number2795480
Product Code NIQ
UDI-Device Identifier07640130439146
UDI-Public7640130439146
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2021
Device Model Number364505
Device Catalogue NumberSEE MODEL NO.
Device Lot Number06193121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Received by Manufacturer 09/03/2020
Supplement Date Received by Manufacturer11/03/2020
Initial Report FDA Received Date09/03/2020
Supplement Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-