Model Number 1581-02-112
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); Pain (1994); Discomfort (2330); No Code Available (3191)
Date of Event 08/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision due to instability.Tibial insert revised.Date of implant: (b)(6) 2019, date of revision: (b)(6) 2020, (left knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2020, patient was revised to address global instability of the left knee.Patient presented with pain, discomfort, recurrent effusion, and global instability.No evidence of loosening or infection was identified.All products were found to be well-fixed intraoperatively.Insert was upsized, tissue releases performed, and stability achieved.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA PLI XLK INS 2.5 12.5MM
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10489808
Report Number1818910-2020-19441
Device Sequence Number16848159
Product Code JWH
UDI-Device Identifier10603295062448
UDI-Public10603295062448
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K040166
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1581-02-112
Device Catalogue Number158102112
Device Lot Number8943702
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2020
Supplement Date Received by Manufacturer09/03/2020
10/05/2020
Initial Report FDA Received Date09/03/2020
Supplement Report FDA Received Date09/16/2020
10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 3 40G; DEPUY CMW 3 40G; PFC*SIGMA C/R NPOR FEM LT SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG MOD TIB TRAY CEM COCR 2.5; DEPUY CMW 3 40G; DEPUY CMW 3 40G; PFC*SIGMA C/R NPOR FEM LT SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG MOD TIB TRAY CEM COCR 2.5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight91
-
-