• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE SPARTACORE GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE SPARTACORE GUIDE WIRE Back to Search Results
Model Number 1005203
Medical Device Problem Codes Break (1069); Difficult to Insert (1316); Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The product risk assessment identifies this as a foreseeable event.The investigation was unable to determine a cause for the reported stretched coil, break and difficult to insert events.It may be possible that during insertion of the wire into the bdc the coils were stretched, possibly due to a break, and that created the resistance making it difficult to insert.This could not be confirmed because the guide wire was not returned.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that after inserting into the balloon dilatation catheter (bdc) with resistance noted, the tip of the spartacore guide wire was shredded/coils were exposed [possible breakage] and unable to be used.The guide wire was immediately removed.It was confirmed that nothing remained in the patient.A new guide wire was used to complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE SPARTACORE GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10490567
Report Number2024168-2020-07370
Device Sequence Number11434808
Product Code DQX
UDI-Device Identifier08717648014031
UDI-Public08717648014031
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K963702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number1005203
Device Catalogue Number1005203
Device Lot Number9121261
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2020
Initial Report FDA Received Date09/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-