| Model Number |
11-4741 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a procedure to install a 10-level pedicle screw construct, the surgeon reported a single screw separated during attempted placement into the ilium.The screw was partially removed and the construct completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The removed portion of the device was returned and the reported event was confirmed.The returned device displayed visual indications of use in a manner that exceeded the strength of the assembly.A review of the dhr found no nonconformances.Based on the results of the evaluation a definitive conclusion cannot be made.
|
| |
|