MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP XLPE INSERT 49/28 NON-CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP XLPE INSERT 49/28 NON-CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 75018956
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 08/04/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tha surgery had been performed, the patient experienced a mechanical dislocation of the ceramic ceramic ball head delta 28 m and polarcup xlpe insert 49/28 non-cem.A revision surgery was performed to replace such devices.The patient outcome is unknown.
 
Additional Manufacturer Narrative
It was reported that, after a tha surgery had been performed, the patient experienced a mechanical dislocation of the ceramic ceramic ball head delta 28 m (75007452) and polarcup xlpe insert 49/28 non-cem (75018956).Both devices used in treatment were received for investigation.A visual inspection was conducted.The polarcup xlpe insert and the ceramic ball head are still connected, indicating that the reported dislocation happened between the insert and shell.The polarcup xlpe insert shows three large dents on the pole of the outer side of the insert.These might origin from impact during dislocation, but also have happened during replacement procedure.Apart from that no obvious damage, such as dents or wear marks, could be observed on the insert.The insert and the ball head were still running smoothly.The size of the polarcup xlpe insert was verified with the "size identification drawing".Size and diameter of the insert are in accordance with the requirements.A medical investigation was performed.Based on the limited information provided, a thorough medical assessment could not be performed.A review of the complaint history revealed no additional complaint for the batch in question.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported dislocation.The ifu (lit.No.12.23 ed 05/16) lists dislocation as a known possible side effect resulting from a hip arthroplasty.A review of the risk management documentation verifies the failure mode and severity of the reported issue.Based on the conducted investigation the reported failure mode cannot completely be confirmed.Without further information about the corresponding shell, the root cause for the dislocation of the xlpe insert from the shell cannot be determined.There is no indication that the device failed to match specification at the time of manufacturing and the risk level of this issue is determined as low.Therefore, the need for corrective action is not indicated.Nevertheless, smith + nephew will continue to monitor this device for similar issues.The returned device will be retained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP XLPE INSERT 49/28 NON-CEM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key10490832
Report Number9613369-2020-00163
Device Sequence Number19504957
Product Code LPH
UDI-Device Identifier07611996111460
UDI-Public07611996111460
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K110135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number75018956
Device Catalogue Number75018956
Device Lot NumberB2004282
Initial Date Received by Manufacturer 08/06/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date09/03/2020
Supplement Report FDA Received Date10/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-