|
Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Device Tipped Over (2589)
|
Patient Problems
Pain (1994); Perforation of Vessels (2135); Ambulation Difficulties (2544)
|
Event Type
Injury
|
Manufacturer Narrative
|
Occupation: non-healthcare professional.Investigation: the reported allegations have been investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation and tilt.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Catalog number and lot number are unknown.The alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
The following information is alleged: the patient received a gunther tulip on (b)(6) 2009.Approximately 9 years and 3 months later, the patient had a computed tomography (ct) scan which revealed that the filter had tilted and perforated beyond the inferior vena cava (ivc) lumen.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Additional information: investigation: the following allegations have been investigated: pain, limited mobility.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported pain and limited mobility is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot number are unknown.The alleged tulip is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2009 due to pulmonary embolism.Patient is alleging tilt, vena cava perforation.Patient further alleges pain and limited mobility.Report from computerized tomography (ct): " lnfrarenal ivc filter is noted.There is apparent extension of the distal aspect of the legs beyond the ivc lumen is best seen on image #41.19.3 [degrees] of medial tilt is noted.No atheromatous calcifications of the aorta or branch vessels noted.".
|
|
Search Alerts/Recalls
|
|
|