JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT525
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to (b)(4) has been submitted.
 
Event or Problem Description
Customer reported that the wrong power was initially implanted in the patient¿s right eye.The iol was explanted and replaced with model zcu450 17.5 diopter power.There was no additional information provided.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 08/25/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed what appeared to be viscoelastic on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10491255
Report Number9614546-2020-00367
Device Sequence Number14095131
Product Code HQL
UDI-Device Identifier05050474603516
UDI-Public(01)05050474603516(17)200831
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberZCT525
Device Catalogue NumberZCT525U170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer09/07/2020
10/25/2020
Initial Report FDA Received Date09/03/2020
Supplement Report FDA Received Date09/29/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Date Report Sent to FDA01/01/2005
-
-