MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; STENT, ILIAC Back to Search Results
Model Number BXA061902A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event or Problem Description
On (b)(6) 2020, the patient presented with an unknown etiology in the renal artery and underwent treatment using a gore® viabahn® endoprosthesis (vsx) and three gore® viabahn® vbx balloon expandable endoprosthesis (vbx).The physician deployed the first vbx device approximately <1 cm shallow in the treatment zone.Then the physician extended the placed endoprosthesis with an additional vbx device.The physician then deployed a vsx device to line both vbx devices.While attempting to remove the rosen guidewire, the wire dislodged all of the stents and pulled the vbx & vsx devices out of the vessel.The physician was then able to snare all three devices and capture the devices for removal with a large bore sheath.A femoral cut down was then used to remove all snared devices from patient.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
H6.Codes updated to reflect completed investigation: investigation findings: code 213.Investigation conclusions: code 4315.Patient information was made available and the following fields were updated: a2.Dob: (b)(6) 1937; 83 years old.A3.Female patient information was made available and the following field was corrected.A1.Patient identifier corrected to a.R.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
MDR Report Key10491595
Report Number2017233-2020-01216
Device Sequence Number17137443
Product Code PRL
UDI-Device Identifier00733132637171
UDI-Public00733132637171
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/10/2022
Device Model NumberBXA061902A
Device Catalogue NumberBXA061902A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/06/2020
Supplement Date Received by Manufacturer08/06/2020
Initial Report FDA Received Date09/03/2020
Supplement Report FDA Received Date10/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
-
-