| Model Number |
MC2-4570S |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for investigation as it is in situ.Root cause is unknown at this time.
|
| |
|
Event or Problem Description
|
|
Information was received that at the first lengthening, x-ray images revealed that there was migration of the end cap from the actuator housing.The patent will be monitored at future lengthening clinics.
|
| |
|
Additional Manufacturer Narrative
|
|
Even though the device remains implanted and no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
|
| |
|