CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY, CLASSIC, 135 CM; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number DBEC-125
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Myocardial Infarction (1969); Thrombus (2101); Injury (2348)
Date of Event 07/14/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Under lidocaine 1% local anesthesia a 6 fr introducer was placed in the right radial artery using modified seldinger technique.The circumflex into om lesion was crossed using a 300 cm length pt moderate support wire.An attempt to advance a 1.25 otw balloon was not successful.A fine cross catheter was used to advance beyond the blockage.The long pt wire was exchanged to viper wire.The 1.25 tip orbital atherectomy device (csi) was advanced using the usual technique.Low speed orbital atherectomy was done at the proximal lesion, followed by distal om lesion.The burr advanced antegrade into the lesion.Upon performing the csi from the distal end into the proximal end of the distal lesion using low speed rpm, the device tip separated from shaft.Approximately 3-4 cm of the distal device broke off.A second wire was used to rewire the lesion.Under lidocaine 1% local anesthesia a 6 fr introducer was placed in the right femoral artery using modified seldinger technique.Several attempts were made to pull back the tip into the guiding catheter but were unsuccessful.This included inflating the balloon distal to the tip and pulling it back.Also included using balloon with ntg proximal to the tip and trying to pull back.Attempted to place a snare, but was unsuccessful.Patient eventually thrombosed the circumflex vessel.Wire and diamond back tip remain in the circumflex.Surgery was consulted but patient was not a surgical candidate.Under lidocaine 1% local anesthesia a 9 fr introducer was placed in the left femoral artery using modified seldinger technique.Iabp was inserted in the left femoral area.Impella cp was inserted in the right femoral area.Iabp removed from left femoral site.Full anti-coagulation started.Tr band deployed to right radial artery.10 ml air in band.Site stable.Right femoral sheath remains in place with impella device in place.Site stable.Left femoral sheath remains in place.Site stable.The patient suffered another myocardial infarction in the heart tissue fed by the left circumflex coronary artery and suffered acute kidney injury and profound hypotension the evening of the procedure.Patient discharged with ongoing care by cardiology.Note: when the device was retrieved from cath lab, all the parts were present but a foreign object was also found in the wrappings and a photo is included in this report.It is unknown if this was part of the device or something else.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY, CLASSIC, 135 CM
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key10493421
Report Number10493421
Device Sequence Number16944763
Product Code MCX
UDI-Device Identifier10852528005787
UDI-Public(01)10852528005787(17)220131(10)307433
Combination Product (Y/N)N
Initial Reporter StateIA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/03/2020
Report Date (Section F) 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberDBEC-125
Device Catalogue Number70058-13
Device Lot Number307433
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Device Age0 YR
Event Location Hospital
Date Report to Manufacturer09/04/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight117
Date Report Sent to FDA09/03/2020
-
-