Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 37092, serial#: unknown, product type: accessory.Product id: 37092, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient programmer (pp) does not communicate/connect to implanted neurostimulator for the last few days.Pp shows 'sync now' screen and does not sync.Pt replaced batteries 3 times and they are alkaline.Pt also noticed in the last few days that the pp antenna is loose in the opening.Replacement requests were processed.No symptoms were reported.
|
| |
|
Event or Problem Description
|
|
Additional information was received from the patient stating that the cause of or not being able to connect to the implant was because the controller was broken due to normal use.A replacement controller was sent which resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d information references the main component of the system and other applicable components are: product id: 37092, serial# unknown, product type: accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the programmer (s/n (b)(6) ) found the digital board was corroded.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
(b)(6) 2020:patltr (con): additional information was received from the patient stating that the cause ofor not being able to connect to the implant was because the controller was broken due to normal use.A replacement controller was sent which resolved the issue.(b)(6) 2020 interface update details: no new information.
|
| |
|