• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE METER Back to Search Results
Model Number 9697
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
An advocate reported that the customer's blood glucose readings from (b)(6) 2020 were not stored in the memory of the contour next meter.During the call, the customer performed a blood glucose test and the reading was properly stored in the meter's memory.There was no allegation of an adverse event.The advocate was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next meter for evaluation.No test strips were returned.Therefore, the returned meter was tested with the in-house contour next test strips using a blood sample, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla, ny
MDR Report Key10495129
Report Number1810909-2020-00445
Device Sequence Number14095144
Product Code NBW
UDI-Device Identifier00301939697010
UDI-Public00301939697010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121190
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9697
Device Catalogue Number9697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer11/11/2020
Initial Report FDA Received Date09/04/2020
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-