| Model Number |
9697 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
An advocate reported that the customer's blood glucose readings since (b)(6) 2020 were not being stored in the memory of the contour next meter.There was no allegation of an adverse event.The advocate was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next meter for evaluation.The returned meter was tested with the in-house contour next test strips using a blood sample, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
|
| |
|