ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550250-38
Medical Device Problem Codes No Apparent Adverse Event (3189); Activation Failure (3270)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: estimated date.The additional xience sierra device referenced is filed under a separate medwatch report number.The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the 3.5x38mm and 2.50x38mm xience sierra stents delivery system failed to deploy.Both devices were removed, and another unspecified stent was used to successfully complete the procedure.There was no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.General handling damage was noted during return analysis consistent with operational context; however, as no specific reported device malfunction was reported nothing was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.An investigation was not conducted because there were no specific reported device malfunctions or patient effects associated with the device.There is no indication of a product quality issue with respect to design, manufacture or labeling.H6: device code 3270 removed.
 
Event or Problem Description
Subsequent to the initially filed report, the following information was received: the devices were not deployed as it was noted that the incorrect size stents were chosen and they were undersized for the lesion.The stents were not implanted and there was no issue noted with the stents or delivery systems.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10495299
Report Number2024168-2020-07405
Device Sequence Number3663181
Product Code NIQ
UDI-Device Identifier08717648227141
UDI-Public08717648227141
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/23/2021
Device Model Number1550250-38
Device Catalogue Number1550250-38
Device Lot Number0031141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Initial Date Received by Manufacturer 08/14/2020
Supplement Date Received by Manufacturer10/28/2020
Initial Report FDA Received Date09/04/2020
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-