| Model Number |
FRED3536-PMA |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
08/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for evaluation.The root cause cannot be determined.The instructions for use (ifu) identifies vessel stenosis or thrombosis as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the fred was used as treatment for a tortuous m2 to m1 fusiform aneurysm with two curves.The fred was delivered smoothly; however, the proximal stent did not completely open upon deployment.Thrombus formation was observed inside the proximal portion of the fred.Integrilin was administered through the microcatheter at the proximal portion of the device.Aspiration was attempted, but was not successful.After multiple manipulations of the device with the microcatheter and microwire, the fred completely opened and the thrombus within the device was no longer visualized.No neurological deficits or abnormalities were observed during the post-procedure neurological exam of the patient.The patient is currently reported to be doing well.
|
| |
|