| Model Number |
2341-1031S |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device was implanted, only packaging is available.
|
| |
|
Event or Problem Description
|
|
As reported: "there was a foreign material like hair between film and packaging." "the nail was used for procedure.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since only the packaging was received unpacked and no other evidence was provided [e.G.Image of packaging pollution in intact and unpacked condition].More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.The presence of the foreign material could have been best presented in original condition, which would have been possible in this case.Nothing was reported during incoming inspection at distribution site respectively during incoming inspection at the hospital itself or before surgery.The foreign material is obviously and would have been detected during inspection.However, as advised per ifu an inspection is recommended prior to surgery.It could not be excluded that the pollution occurred during storage/handling at user site.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.
|
| |
|
Event or Problem Description
|
|
As reported: "there was a foreign material like hair between film and packaging." "the nail was used for procedure.".
|
| |
|