• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH EXTRACTOR ASNIS III FOR SCREWS:4.0MM AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH EXTRACTOR ASNIS III FOR SCREWS:4.0MM AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 702496
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "the tip of the extractor was damaged during removal." damage observed: breakage.
 
Event or Problem Description
As reported: "the tip of the extractor was damaged during removal." damage observed: breakage.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received extractor shows signs of normal and prolonged usage however its tip is found to be broken.The breakage surface shows mixture of rough and smooth surface.Absence of plastic deformation indicates that the breakage was instantaneous due to high torsional and bending overload.However, it cannot be excluded that residual stress from previous usage also contributed in the breakage.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is user related.The breakage of the tip happened due to a torsional and bending overload, the likelihood of which is greater during an explantation.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR ASNIS III FOR SCREWS:4.0MM AO FITTING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10496897
Report Number0008031020-2020-02215
Device Sequence Number2147099
Product Code LXH
UDI-Device Identifier04546540112415
UDI-Public04546540112415
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number702496
Device Catalogue Number702496
Device Lot Number12416D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Received by Manufacturer 08/12/2020
Supplement Date Received by Manufacturer10/15/2020
Initial Report FDA Received Date09/04/2020
Supplement Report FDA Received Date11/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-