| Model Number |
702496 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the tip of the extractor was damaged during removal." damage observed: breakage.
|
| |
|
Event or Problem Description
|
|
As reported: "the tip of the extractor was damaged during removal." damage observed: breakage.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received extractor shows signs of normal and prolonged usage however its tip is found to be broken.The breakage surface shows mixture of rough and smooth surface.Absence of plastic deformation indicates that the breakage was instantaneous due to high torsional and bending overload.However, it cannot be excluded that residual stress from previous usage also contributed in the breakage.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is user related.The breakage of the tip happened due to a torsional and bending overload, the likelihood of which is greater during an explantation.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Search Alerts/Recalls
|