PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY Back to Search Results
Model Number 5MAXACE068
Medical Device Problem Codes Break (1069); Suction Problem (2170)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the distal internal carotid artery (ica) using a penumbra system ace 68 reperfusion catheter (ace68), a neuron max 6f 088 long sheath (neuron max), and a guidewire.During the procedure, the physician inserted the neuron max and ace68 into the patient, then connected a syringe to the ace68; however, no aspiration was observed.Subsequently, the physician realized that there was no negative pressure.The physician then made another attempt, but same issue occurred.Therefore, the ace68 was removed and flushed on the back table.While flushing the ace68, the physician found a leak approximately 1-2 centimeters near the distal tip.Therefore, the ace68 was no longer used.The procedure was completed using a new ace68, the same neuron max, and the same guidewire.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10497243
Report Number3005168196-2020-01393
Device Sequence Number2687548
Product Code NRY
UDI-Device Identifier00814548016535
UDI-Public00814548016535
Combination Product (Y/N)Y
Initial Reporter CountryTW
PMA/510(K) Number
K161640
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/13/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/29/2022
Device Model Number5MAXACE068
Device Catalogue Number5MAXACE068
Device Lot NumberF92102
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2020
Initial Report FDA Received Date09/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-