• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10607
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/31/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The 75% stenosed target lesion was located in the mildly tortuous and mildly calcified mid to distal right coronary artery.A 4.00 x 38 synergy drug-eluting stent was selected for use.However, upon removing the stylet and sheath, resistance was felt and it was noted that the stent was damaged.There were missing struts in the mid to distal section of the stent.The procedure was completed with another of the same device.
 
Event or Problem Description
It was reported that stent damage occurred.The 75% stenosed target lesion was located in the mildly tortuous and mildly calcified mid to distal right coronary artery.A 4.00 x 38 synergy drug-eluting stent was selected for use.However, upon removing the stylet and sheath, resistance was felt and it was noted that the stent was damaged.There were missing struts in the mid to distal section of the stent.The procedure was completed with another of the same device.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy ii us mr 4.00 x 38 stent delivery system catheter was returned for analysis.A visual examination of the stent found stent damage.Stent struts on the 5th and 6th proximal rows were lifted and bunched distally.All stent struts were present.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10497659
Report Number2134265-2020-12349
Device Sequence Number14114150
Product Code NIQ
UDI-Device Identifier08714729840435
UDI-Public08714729840435
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2021
Device Model Number10607
Device Catalogue Number10607
Device Lot Number0024666504
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2020
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer10/29/2020
Initial Report FDA Received Date09/04/2020
Supplement Report FDA Received Date11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-