• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE CERTAIN IMPLANT 4 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE CERTAIN IMPLANT 4 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number IOSS410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 01/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).Patient identifier unknown / not provided.Weight unknown / not provided.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.A summary investigation has also been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported implant removed due to loss of integration with peri-implantitis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE CERTAIN IMPLANT 4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10498678
Report Number0001038806-2020-01308
Device Sequence Number11503728
Product Code DZE
UDI-Device Identifier00844868007081
UDI-Public(01)00844868007081(17)130303(10)2007120350(241)IOSS410
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/03/2013
Device Model NumberIOSS410
Device Catalogue NumberIOSS410
Device Lot Number2007120350
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2020
Initial Report FDA Received Date09/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
-
-