• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC NATURA; PROTECTOR, OSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC NATURA; PROTECTOR, OSTOMY Back to Search Results
Model Number 413419
Medical Device Problem Codes Malposition of Device (2616); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device 7 of 10.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
The end user stated that in 10 out of 10 wafers, the stoma opening was off-centered.He had used 8 out of 10 wafers and this issue had resulted in a decreased wear time of 2-3 days.Reportedly, his usual wear time was 5 days.There was no harm reported and no photo is available at this time.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi).H6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions.H10: investigation summary.Returned sample evaluation: there are no photographs associated with this case no unused returned sample was received.Conclusion summary of the related event: a complaint search performed in database by manufacturing date for the ¿starter hole off center or misaligned¿ failure mode for ats#1 and ats#2 manufacturing lines shows a decreasing trend from jun 2019 to date (cut off september 30, 2020).Based on the preliminary investigation, the most probable cause for the cases related to manufacturing process in the ats#1 and ats#2 manufacturing lines is related to machine condition.As part of the nonconformance report (ncr), actions were taken for this failure mode, which were the main factors for the reduction in the quantity of complaints per month.In addition, the root cause investigations (rci¿s) and were carried out and actions were taken for the off-center failure mode in the convex two (2) piece (ostomy) and convex two (2) piece (wct) bulks manufacturing lines.Furthermore, a nonconformance search was performed in database from 2018 to september 30, 2020, for the ¿starter hole off center or misaligned¿ failure mode for ats#1 and ats#2 manufacturing lines and as a result, no events related to this malfunction were found.Moreover, the following quality and manufacturing controls are currently in place in the ats#1 & ats#2, according to the product instructions and test methods, which contribute directly to the detection activities: manufacturing process instructions (pi21-130 - ats1 assembly and packaging / pi21-122 - ast2 assembly and packaging).1.Visual inspection to completed product 100% of units.2.Testing and inspection log ¿ dimensional verification and visual inspection looking for collar concentricity 6 samples at the beginning of the batch and each 1 hour.Quality test methods (tm-017 ¿ visual nonconformities laminated adhesive products / tm-014 ¿ pouch nonconformities).A quality notification was given to the quality and manufacturing personnel of the ats#1, ats#2, convex two (2) piece (ostomy) and convex two (2) piece (wct) manufacturing lines, to make them aware of the new complaints received for this failure mode (refer to attachment #3 and attachment #5 of this ncr).For everything explained above, no additional actions are required.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURA
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key10499200
Report Number9618003-2020-13527
Device Sequence Number14243385
Product Code EXE
UDI-Device Identifier00768455124015
UDI-Public00768455124015
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2024
Device Model Number413419
Device Lot Number9K02485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/14/2020
Supplement Date Received by Manufacturer11/20/2023
Initial Report FDA Received Date09/07/2020
Supplement Report FDA Received Date11/24/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-