Model Number 6260-9-244
Medical Device Problem Codes Material Erosion (1214); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Injury (2348); Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
Date of Event 08/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation is being performed in an attempt to identify the cause of the event.Should additional information become available it will be reported in a supplemental report.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Device not returned.
 
Event or Problem Description
Patient came to (b)(6) clinic complaining of progressive hip pain.An x-ray revealed a disassociated femoral head due to excessive trunion wear of the femoral stem.During surgery was noted the trunion had worn to the point that the femoral head had came off.After explanting the accolade tmzf stem a restoration modular stem was used as the replacement device.
 
Event or Problem Description
Patient came to cleveland clinic complaining of progressive hip pain.An x-ray revealed a disassociated femoral head due to excessive trunion wear of the femoral stem.During surgery was noted the trunion had worn to the point that the femoral head had came off.After explanting the accolade tmzf stem a restoration modular stem was used as the replacement device.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and wear involving a metal head was reported.The event was confirmed via clinician review.Method & results: product evaluation and results: the metal head was not returned and no photo was provided.However, a photo of the mated stem was provided.There are scratch marks and tissue on the explanted stem.The stem trunnion was penciled.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: the event can be confirmed.The photo and the x-ray represented typical findings for failure at the taper/trunnion wear.The event likely represents failure of a recalled lfit device.Obtaining the catalog number and/or the implants may allow confirmation.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical records by a clinical consultant revealed that the event can be confirmed.The photo and the x-ray represented typical findings for failure at the taper/trunnion wear.The subject device has been identified to be within scope of an nc and capa.Lot specific voluntary recall ra 2016-028 was initiated for the lfit v40 cocr heads within scope of this nc capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.Corrective action/preventive action: an nc was initiated on 27-january-2016 due to the occurrence rate for taper lock failure in the risk management files for specific lots of lfit v40 cocr heads.Given the potential risk of nine (9) hazards identified in the hazards and harm evaluation, voluntary product recall ra 2016-028 was issued.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation, wear/metallosis and osteolysis involving a metal head was reported.The event was confirmed via clinician review.Method & results: product evaluation and results: the metal head was not returned and no photo was provided.However, a photo of the mated stem was provided.There are scratch marks and tissue on the explanted stem.The stem trunnion was penciled.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: a review of the provided medical records by a clinical consultant stated the following: adverse event identified: revision of left hip arthroplasty due to disassociation of a femoral head from a tmzf stem and trunnionosis.Hazards: failure of trunnion-head construct.Conclusion of assessment: review of the records provided confirmed the implant disassociation and trunnion wear.Review of implant stickers could define if the head was an lfit or associated with an lfit problem.Obtaining the implant may provide additional insight into the mechanism of disassociation.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical records by a clinical consultant revealed that the event can be confirmed.The photo and the x-ray represented typical findings for failure at the taper/trunnion wear.The subject device has been identified to be within scope of an nc and capa.Lot specific voluntary recall ra 2016-028 was initiated for the lfit v40 cocr heads within scope of this nc and capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.
 
Event or Problem Description
Patient came to (b)(6) clinic complaining of progressive hip pain.An x-ray revealed a disassociated femoral head due to excessive trunion wear of the femoral stem.During surgery was noted the trunion had worn to the point that the femoral head had came off.After explanting the accolade tmzf stem a restoration modular stem was used as the replacement device.Update on 04-aug-2021 per medical review: ¿post-operatively it was stated that there was catastrophic failure of the hip, trunnion failure at the taper, metal disease, femoral osteolysis with a well fixed stem, wear through and destruction of the acetabulum, periacetabular osteolysis and abductor destruction.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 44MM/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10499267
Report Number0002249697-2020-01801
Device Sequence Number11503547
Product Code JDI
UDI-Device Identifier07613327032758
UDI-Public07613327032758
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2012
Device Model Number6260-9-244
Device Catalogue Number6260-9-244
Device Lot NumberP0RMPD
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/11/2020
Supplement Date Received by Manufacturer10/20/2020
08/04/2021
Initial Report FDA Received Date09/07/2020
Supplement Report FDA Received Date11/13/2020
08/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-08/29/2016-007R
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient Weight94
-
-