MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550400-23
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing non-conformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.Factors that may contribute to stent dislodgement include, but not limited to; improper or inadequate crimping at the time of manufacture, incorrect sheath sizing, positive pressure during preparation, forced sheath removal, handling of the stent during preparation, and interaction with the lesion, accessory devices, and/or previously implanted stents.To ensure this is not the result of a manufacturing issue, all stent delivery systems are inspected for proper stent placement, stent damage and crimped stent outer diameter.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that prior to use in the patient and during device preparation, the xience sierra stent dislodged from the balloon.There was no difficulty removing the protective sheath/stylet and no damage to the outer or inner packaging.There was no patient involvement, and no clinically significant delay in the procedure.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10500296
Report Number2024168-2020-07442
Device Sequence Number14258567
Product Code NIQ
UDI-Device Identifier08717648227516
UDI-Public08717648227516
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2020
Device Model Number1550400-23
Device Catalogue Number1550400-23
Device Lot Number9110741
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2020
Initial Report FDA Received Date09/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-